List of services dedicated to laboratory centrifuges.
What strategy should be adopted?
Maintenance services & performance monitoring of laboratory centrifuge equipment:
In order to implement an effective strategy for managing Maintenance & Metrology services associated with the "centrifuge" equipment and to avoid any confusion, it is important to define each of the concepts:
Corrective maintenance Troubleshooting operations may impact equipment performance. Adjustments and performance (metrological) checks may be necessary.
Preventive maintenance Technical inspection of the equipment, mechanical adjustments and lubrication, periodic preventive replacement of wear parts on a case-by-case basis (cylinders, silent blocks, etc.). Objective: to guarantee the long-term performance of the equipment and limit the risk of breakdowns.
Regulatory safety check This applies to centrifuges with a basket (tub) diameter greater than 400 mm if the kinetic energy involved exceeds 1500 joules (salad spinners, in particular, are excluded). The inspection involves verifying the proper functioning of the imbalance detector and the various safety devices. This annual inspection is mandatory as required by Decree No. 2008-244 of March 7, 2008.
Time calibration: Classified under COFRAC calibration activities, this operation involves comparing the cycle time recorded by the equipment to a measurement taken by a reference standard. It is necessary to determine beforehand whether the countdown begins from the start command or when the speed setpoint is reached. This information must be specified on the certificate.
Frequency calibration: Classified under COFRAC calibration activities, this operation involves comparing the speed displayed by the equipment to a measurement taken by a reference standard. The unit of measurement is the number of rotations per minute (RPM). The calculation of the relative centrifugal acceleration (RCA, unit "g") can be provided by the service provider. A radius measurement must be performed (at a minimum to verify the values mentioned in the technical specifications) to obtain this result.
Steady-state temperature test: classified under COFRAC testing activities. This involves a temperature characterization of the centrifuge's charge volume in dynamic (rotating) phase.
Temperature test under transient regime: This study examines the temperature evolution from preconditioning (steady state), through the loading phase (transient state), to the point where the centrifuge reaches stable operating conditions (steady state). The objective of this study is to validate the preparation process for centrifugation.
Unaccredited temperature tests:
– Measurement of heating temperature: calculation of the temperature difference after-before centrifugation inside water tubes installed in the centrifuge; ;
– Display calibration: comparison of the temperature value indicated by the centrifuge display to a measurement made by a reference chain whose probe is positioned next to the regulating probe; ;
– Measurement of dynamic temperature at a point: recording of temperature evolution in centrifugation using an on-board recorder.
Training and best practices:
– General training: training on risks, safety instructions, maintenance, general rules for using centrifuges; ;
– Good practices: specific operating procedure for the use of centrifuges by the laboratory; ;
– User training: targeted training on the installed equipment; ;
– Maintenance and upkeep: frequency and procedures for maintenance, cleaning and internal upkeep specific to each piece of equipment.
Qualification :
Qualification is a set of physical, documentary, environmental and technical controls carried outeduring the commissioning of equipment to ensure compliance with the specifications of the installed equipment and its accessories, to control the installation (power supplies, support, ambient conditions, etc.), to verify its integrity and performance, to allow the customer to use it safely, to train them in its use, and to inform them about the recommended maintenance and upkeep operations.
Definition of the different qualifications:
– Design Qualification (QC): documented verification that the proposed equipment design is suitable for its intended uses; ;
– Installation Qualification (IQ): documented verification that the equipment, as installed, conforms to the approved design and manufacturer's recommendations: conformity of delivery (checking of delivered items, conformity of order and documentation), verification of expectations (power supply, fluids, etc.), assembly, connection and start-up, inspection, staff training; ;
– Qualification to operational (QO): documented verification that the equipment, as installed, functions as intended across the entire operating range: "non-specific" verification program.) Drafting and issuing a report, control and delivery of attached documents; ;
– Performance qualification (QP): Documented verification that the equipment, as arranged, is capable of operating effectively and reproducibly, based on the approved operating method and product specification: "specific" verification program. Drafting and issuing of a control report and delivery of attached documents.
Final operations: Integration of equipment into the fleet / creation of a lifecycle record / commissioning.
Good to know (and little known): The standard NF EN 61010-2-020 "Safety rules for electrical measuring, control and laboratory equipment/Part 2-020: Particular requirements for laboratory centrifuges" provides for equipment installation procedures which can be verified during periodic checks.
Among other information, the following must appear:
– Identification and list of usable rotors and accessories; ;
– List of restrictions defined by the manufacturer on the materials to be centrifuged or not, the density and volume limits; ;
– Marking and respecting a free space of 30cm around the centrifuge; ;
– Procedure for loading the centrifuge and changing the rotor if necessary; ;
– List of ways to work safely near the centrifuge; ;
– Biological safety instructions: use of casings, etc.
– List and frequency of maintenance points: inspection of housings, rotors, fixings, etc.
– Cleaning and decontamination instructions in case of dispersion of material in the centrifuge; ;
– Manufacturer's recommendations for cleaning / decontamination.
Points to consider when implementing a management strategy for laboratory centrifuge equipment:
Corrective maintenance: Infrequently! Effective preventative maintenance reduces the risk of breakdowns and accidents. This operation should be entrusted to the manufacturer or an authorized service provider. Performance checks should be scheduled if necessary.
Preventive maintenance: Annual frequency or as recommended by the manufacturer. Inspection and replacement of wear parts should be carried out by the manufacturer or an authorized service provider.
Regulatory safety inspection: Mandatory annual testing. It is advisable to select a service provider experienced with this type of equipment (risk of damaging the equipment during imbalance testing). This service can be combined with preventive maintenance of the equipment.
Time-frequency calibration: Calibration should be performed annually or every two to five years using simple monitoring methods (drift monitoring with a stopwatch, speed measurement with a tachometer). No drift in these parameters is observed on the centrifuges due to the design and technological performance of the control systems. The recommended calibration program for time and frequency parameters is: min-average-max for general use, or according to the operating instructions for specific use. Time and frequency calibrations can be combined with preventive maintenance services and regulatory safety checks.
Temperature tests:
– Steady-state temperature test: Annual, biennial to quinquennial frequency with implementation of monitoring methods (display drift tracking by comparison with a reference probe, dynamic temperature measurement at a single point). Effective preventive maintenance helps limit the risk of equipment drift (maintenance of production and temperature control systems). Recommended test program: all speed-temperature combinations (impact of speed on temperature control).
– Temperature test under transient regime: During centrifuge installation to validate the pre-conditioning process for the various instructions. Recommended test program: operating instructions.
It is possible to combine temperature tests with time-frequency calibrations, preventive maintenance services, and regulatory safety checks.
Training and best practices:
– General training: training provided by a specialized organization enabling the risk prevention approach, the development of specifications, the implementation of an appropriate maintenance plan; ;
– Good practices: internal training provided by a competent reference staff member (general training + regulatory context specific to the sector of activity), keep the evidence; ;
– User training: training to be provided during equipment installation and before its authorization to be put into production, keep the evidence; ;
– Maintenance and upkeep: setting up lists of actions to be carried out at defined frequencies according to the manufacturer's recommendations and the use of the equipment by the laboratory, keeping the evidence.
Qualifications (optional):
– Design Qualification (QC): applicable to the design of "custom-made" equipment; ;
– Installation Qualification (IQ): Prepare the specifications for the equipment and its accessories, determine the installation conditions (power supplies, mounting hardware, ambient conditions), and ascertain the supplier's capacity to provide qualification services. This is a responsibility specific to the equipment manufacturer or supplier. Installation options are available in accordance with standard NF EN 61010-2-020.
– Qualification to operational QO/QP: Provide a review of metrological requirements, programming of equipment to be installed. Calibration and testing operations for the various quantities for specific (OQ) and/or non-specific (PQ) use.
Example of a management strategy for "laboratory centrifuge" equipment:
Corrective maintenance: entrusted to the after-sales service of the authorized equipment distributor (to be verified beforehand); plan for the necessary metrological checks in the event of replacement of component(s) involved in regulating a parameter of the equipment (speed control, time control, refrigeration intervention, etc.). Rely on the expertise of the after-sales service to assess the need to verify the equipment's performance following the intervention.
Preventive maintenance: entrusted to the after-sales service of the authorized equipment distributor (to be checked beforehand).
Regulatory safety inspection: annual, entrusted to the after-sales service of the equipment distributor and coupled with preventive maintenance.
Time-frequency calibration: Option 1: Calibration under accreditation every five years with annual internal verification using a stopwatch and an uncalibrated tachometer. Periodic verification of the instruments by comparing the indicated values with the calibration results of a centrifuge (used as a generator). Option 2: Calibration under accreditation every year, combined with preventive maintenance and regulatory safety checks. Calibration program in terms of time and frequency: guidelines used. Calibrations entrusted to the accredited metrology laboratory responsible for the metrological management of the equipment.
Temperature tests:
– Temperature tests under steady-state conditions: Tests under accreditation at a five-year frequency with implementation of annual monitoring at one point using the dynamic temperature measurement method. Test program: each speed-temperature pair used.
– Temperature tests under transient regime: Installation of the centrifuge. Test program: instructions used.
Tests entrusted to the accredited metrology laboratory in charge of the metrological management of equipment.
Training and best practices:
– General training: external training for the safety manager provided by a competent body; ;
– Good practices: training provided internally by the quality department (use of crankcases, balancing methods, etc.);
– User training: training on the use of the equipment provided during installation by the supplier (programming, maintenance, etc.); ;
– Maintenance and upkeep: list, products, frequency and validation of cleaning, decontamination (tank, rotor, etc.) and periodic maintenance operations (anti-corrosion protocol, condenser dust removal, etc.).
Qualifications:
– Installation Qualification (IQ): verification of expectations, environmental conditions, delivery conformity, associated documents, specific training on use, submission of a qualification file by the supplier; ;
– Qualification to operational QO/QP: recording and verification of the various customer programs (QP), handing over of traceability evidence to the SI (if metrological services are outside accreditation), handing over of a qualification file by the metrology laboratory of the equipment supplier accredited for the different quantities (time-frequency-temperature).
And tomorrow?
IoT (connected sensors) are expected to soon appear in the field of centrifugation. Their contribution? Continuous monitoring of physical quantities, the proper functioning of safety and control devices, and mechanical vibrations, in order to enable predictive and efficient maintenance.